Choice Medical MD300C - Possible Problems and Resolutions

 

email bookmark digg it

 

Choice Medical MD300C Fingertip Pulse Oximeter - Problems and Resolutions

Possible Problems and Resolutions

Problem: Spo2 or PR can not be shown normally

 

Possible reason: Finger is not plugged correctly

Solution: Retry by plugging the finger

 

Possible reason: Patient's Oxyhemoglobin value is too low to be measured

Solution: Try some more times. If you can make sure about no problem existing in the product, please go to a hospital for exact diagnosis

 

Possible reason: Finger is not plugged correctly

Solution: Retry by plugging the finger

 

Problem: Spo2 or PR is shown unstably

 

Possible reason: Finger might not be plugged deep enough

Solution: Retry by plugging the finger

 

Possible reason: Finger is trembling or patient's body is in movement status

Solution: Try not to move

 

Problem: The Oximeter cannot be powered on

 

Possible reason: Powerof batteries might be inadequate or not be there at all

Solution: Replace batteries

 

Possible reason: Batteries might be installed incorrectly

Solution: Reinstall the batteries

 

Possible reason: The Oximeter might be damaged

Solution: Contact local customer service centre

 

Problem: Indication lamps are suddenly off

 

Possible reason: The product is automatically powered off when no signal is detected longer than 8 seconds

Solution: Normal

 

Possible reason: Low power

Solution: Replace the batteries

 

Problem: "Error3" or "Error4" displayed on screen

 

Possible reason: Low power

Solution: Change new battery

 

Possible reason: Receiving tube being shielded or damaged together with broken connector.

Solution: Contact local customer service centre

 

Possible reason: mechanical Misplace for receive-emission tube

Solution: Contact local customer service centre

 

Possible reason: amp circuit malfunction.

Solution: Contact local customer service centre

 

Problem: "Error7" displayed on screen

 

Possible reason: Low power

Solution: Change new battery

 

Possible reason: Emission tube damaged.

Solution: Contact local customer service centre

 

Possible reason: Current control circuit malfunction.

Solution: Contact local customer service centre

invisible

 

 

 

alien

secure payments

 

Member Login

Currently Not Logged In

AmperorDirect.com

Menu Bottom

 

Member Login

Your shopping basket is currently empty.

Menu Bottom

 

Member Login

We ship, on a daily basis, packages all over the world!

Our Customers' Locations

 

Member Login

Menu Bottom

 

Member Login

Menu Bottom

GPS Store

Everything Mobile Customers Visit our GPS Store!

  Copyright frame