Choice Medical MD300C - Possible Problems and Resolutions
Choice Medical MD300C Fingertip Pulse Oximeter - Problems and Resolutions
Possible Problems and Resolutions
Problem: Spo2 or PR does not displayed
Possible reason: Finger is not inserted correctly
Solution: Re-insert the finger
Possible reason: Patient's SpO2 value is too low to be measured
Solution: Increase your hands' blood circulation by rubbing them gently toegether for a few minutes. Try a few more times. If you can make sure that no problem exists in the product, please go to a hospital for exact diagnosis
Problem: Spo2 or PR reading is unstably
Possible reason: Finger is not inserted correctly.
Solution: Re-insert the finger
Possible reason: Excessive patient movement
Solution: Restrict patient / oximeter movement.
Possible reason: Patient has heart flutter or atrial fibrillation.
Solution: Try some more times. If you can make sure about no problem existing in the product, please go to a hospital for exact diagnosis
Problem: The Oximeter cannot be powered on
Possible reason: No battery or dead battery
Solution: Replace batteries
Possible reason: Batteries might be installed incorrectly
Solution: Reinstall the batteries
Possible reason: The Oximeter might be damaged
Solution: Contact local customer service centre
Problem: Display suddenly turns off
Possible reason: The oximeter will automatically powered off when no signal is detected in 8 seconds
Solution: Normal
Possible reason: Battery power is low
Solution: Replace batteries
Problem: "Error3" or "Error4" displayed on screen
Possible reason: Battery power is low
Solution: Replace batteries
Possible reason: Receiving tube being shielded or damaged together with broken connector.
Solution: Please contact local customer service centre
Possible reason: Mechanical Misplace for receive-emission tube
Solution: Please contact local customer service centre
Possible reason: Amp circuit malfunction.
Solution: Please contact local customer service centre
Problem: "Error7" displayed on screen
Possible reason: Battery power is low
Solution: Replace batteries
Possible reason: Emission tube damaged.
Solution: Please contact local customer service centre
Possible reason: Current control circuit malfunction.
Solution: Please contact local customer service centre
Error Codes (latest updated from manufacturer)
- Err 1: program memory damaged.
- Err 2: data memory damaged.
- Err 3: sensor Red Emission Tube damaged.
- Err 4: sensor Infrared-ray Emission Tube damaged.
- Err 5: sensor Infrared-ray Receipt Tube damaged.
- Err 6: exterior crystal oscillator damaged.
- Err 7 & Err 8: sensor emission tube or receipt tube damaged.
- Err 9: real time clock damaged.
- Err 10: EEPROM chip damaged.