Choice Medical MD300C - Possible Problems and Resolutions

Choice Medical MD300C Fingertip Pulse Oximeter - Problems and Resolutions
Possible Problems and Resolutions
Problem: Spo2 or PR can not be shown normally
Possible reason: Finger is not plugged correctly
Solution: Retry by plugging the finger
Possible reason: Patient's Oxyhemoglobin value is too low to be measured
Solution: Try some more times. If you can make sure about no problem existing in the product, please go to a hospital for exact diagnosis
Possible reason: Finger is not plugged correctly
Solution: Retry by plugging the finger
Problem: Spo2 or PR is shown unstably
Possible reason: Finger might not be plugged deep enough
Solution: Retry by plugging the finger
Possible reason: Finger is trembling or patient's body is in movement status
Solution: Try not to move
Problem: The Oximeter cannot be powered on
Possible reason: Powerof batteries might be inadequate or not be there at all
Solution: Replace batteries
Possible reason: Batteries might be installed incorrectly
Solution: Reinstall the batteries
Possible reason: The Oximeter might be damaged
Solution: Contact local customer service centre
Problem: Indication lamps are suddenly off
Possible reason: The product is automatically powered off when no signal is detected longer than 8 seconds
Solution: Normal
Possible reason: Low power
Solution: Replace the batteries
Problem: "Error3" or "Error4" displayed on screen
Possible reason: Low power
Solution: Change new battery
Possible reason: Receiving tube being shielded or damaged together with broken connector.
Solution: Contact local customer service centre
Possible reason: mechanical Misplace for receive-emission tube
Solution: Contact local customer service centre
Possible reason: amp circuit malfunction.
Solution: Contact local customer service centre
Problem: "Error7" displayed on screen
Possible reason: Low power
Solution: Change new battery
Possible reason: Emission tube damaged.
Solution: Contact local customer service centre
Possible reason: Current control circuit malfunction.
Solution: Contact local customer service centre
|