Technical Specifications:
Display Type: OLED display
SPO2:
Measurement range:70-99%
Accuracy: ±2% on the stage of 80%-99%; ±3% on the stage of 70%-79%;
Pulse Rate:
Measure range:30-235 BPM
Accuracy: 30 ~ 100BPM ±2 or 101 ~ 235BMP±2%
Power Requirements:
Two AAA alkaline Batteries
Power consumption: Less than 30mA
Low power indication:
Battery Life: Two AAA 1.5V, 800mAh alkaline batteries could be continuously operated as long as 30 hours.
Dimension:
Length: 55mm
Width: 32mm
Height: 28mm
Weight: 30g (including two AAA batteries)
Environment Requirements:
Operation Temperature: 5-40°C
Storage Temperature: -10-40°C
Ambient Humidity: 15%-80% in operation, 10%-80% in storage
Declaration: EMC of this product comply with IEC60601-1-1-2 standard.
Measurement Performance in Low Perfusion Condition: required the test equipment (BIO-TEK INDEX Pulse Oximeter tester) the pulse wave is available without failure when the simulation pulse wave amplitude is at 6%.
Interference Resistance Capacity against Ambient Light: Device work normally when mixed noise produced by BIO-TEK INDEX Pulse Oximeter tester