HOME | COMPANY | PRODUCTS | CONTACT US | 281.664.1209
Directory: Home | Knowledge Base | Product Support |
Choice Medical Fingertip Pulse Oximeter MD300C1 Reference
Choice Medical MD300C1 Fingertip Pulse Oximeter Support
General Description- Precautions
- Technical Specifications
- Product Description
- Operation Instructions
- Installation
- Maintenance
- Problems and Resolutions
Certification
Useful Tips
Instrument performance depends on the pulsatile character of the artery. The measurement would not be considered reliable and accurate if the following conditions are present during measurement:
- The patient is in cardiac arrest or shock
- Venous pulsations
- Cold hand
- Have taken vascular activity medicine
- Anemia
- Carboxyhemoglobin
- Methemoglobin
- Methylene blue
- Indocyanine green
Error Codes (latest updated from manufacturer)
- Err 1: program memory damaged.
- Err 2: data memory damaged.
- Err 3: sensor Red Emission Tube damaged.
- Err 4: sensor Infrared-ray Emission Tube damaged.
- Err 5: sensor Infrared-ray Receipt Tube damaged.
- Err 6: exterior crystal oscillator damaged.
- Err 7 & Err 8: sensor emission tube or receipt tube damaged.
- Err 9: real time clock damaged.
- Err 10: EEPROM chip damaged.
For more information on pulse oximeter, visit our Pulse Oximeter Resource Center.